dc.creator KORKMAZ, İsmail
dc.creator KAYMAZ, İrfan
dc.creator YILDIRIM, Ömer
dc.date 2015-01-04T19:32:11Z
dc.date.accessioned 2019-07-09T11:58:53Z
dc.date.available 2019-07-09T11:58:53Z
dc.identifier http://dergipark.org.tr/sdufenbed/issue/20805/222229
dc.identifier 10.19113/sdufbed.01928
dc.identifier.uri http://acikerisim.sdu.edu.tr/xmlui/handle/123456789/46015
dc.description Total Knee Arthroplasty (TKA) treatment has been widely used for knee joint, having the most complex motion sequences and shape of the body, in case of loss of motion ability such as gonarthrosis. Lots of failures such as loosening, wear and instability have occurred as a result of increasing usage period. Instability is the most commonly seen failure associated with fixation. Especially loosening of tibial component is more common than that of femoral component. Evaluation of bone in terms of shape related with patient’s weight, height and gender is required to increase the success of TKAs comprising traditional tibial components. Most of the current commercially available knee prostheses are designed based on the anthropometric data of Caucasian knees. This means an accordance problem for the people in different regions having different joint size. In this study, a patient specific tibial component has been designed according to cutting surface of tibial bone obtained from the patient’s computed tomography images. This proposed design has been comparatively investigated with the traditional cemented tibial component by using the finite element analysis. In conclusion, more uniform stress distribution on the tibial trabecular bone related with stress shielding due to long term use was obtained in the patient specific prosthesis. The results have shown that the compatibility
dc.description Total Knee Arthroplasty (TKA) treatment has been widely used forknee joint, having the most complex motion sequences and shape of the body, incase of loss of motion ability such as gonarthrosis. Lots of failures such asloosening, wear and instability have occurred as a result of increasing usageperiod. Instability is the most commonly seen failure associated with fixation.Especially loosening of tibial component is more common than that of femoralcomponent. Evaluation of bone in terms of shape related with patient's weight,height and gender is required to increase the success of TKAs comprisingtraditional tibial components. Most of the current commercially available kneeprostheses are designed based on the anthropometric data of Caucasian knees.This means an accordance problem for the people in different regions havingdifferent joint size. In this study, a patient specific tibial component has beendesigned according to cutting surface of tibial bone obtained from the patient'scomputed tomography images. This proposed design has been comparativelyinvestigated with the traditional cemented tibial component by using the finiteelement analysis. In conclusion, more uniform stress distribution on the tibialtrabecular bone related with stress shielding due to long term use was obtainedin the patient specific prosthesis. The results have shown that the compatibility between the tibial bone and its component is better than the traditional modeland possibility of loosening failure in the patient specific model is lower.
dc.format application/pdf
dc.language en
dc.publisher Süleyman Demirel University
dc.publisher Süleyman Demirel Üniversitesi
dc.relation http://dergipark.org.tr/download/article-file/194022
dc.source Volume: 18, Issue: 3 92-101 en-US
dc.source 1308-6529
dc.title Kişiye Özel ve Geleneksel Tibial Diz Protezi Bileşenlerinin Performans Analizi en-US