dc.creator AKKURT, İskender
dc.creator ARDA, S. S.
dc.creator Gunoglu, Kadir
dc.date 2013-12-31T22:00:00Z
dc.date.accessioned 2020-10-06T09:44:02Z
dc.date.available 2020-10-06T09:44:02Z
dc.identifier 38fbc31b-fdba-4892-82fd-6a2e0011ff3b
dc.identifier 10.1155/2014/186798
dc.identifier https://avesis.sdu.edu.tr/publication/details/38fbc31b-fdba-4892-82fd-6a2e0011ff3b/oai
dc.identifier.uri http://acikerisim.sdu.edu.tr/xmlui/handle/123456789/57572
dc.description As it is important to obtain accurate analytical result in an experimental research, this required quality control of the experimental system. Gamma spectrometry system can be used in a variety of different fields such as radiation and medical physics. In this paper the absolute efficiency, peak to valley ratio, and energy resolution of a 3 '' x 3 '' NaI(Tl) detector were determined experimentally for 511, 662, 835, 1173, 1275, and 1332 keV photon energies obtained from Na-22, Mn-54, Co-60, and Cs-137 radioactive sources.