Purpose: Locking screw deformation is common in nailingof comminuted tibia fractures. The aimis to compare five different proximal locking screws bending resistance intibia nailing system.Methods: 50 screws were divided intofive groups. A stainless steel tube which has a 34 mm internal diameter wasprepared representing the proximal tibia. A 3-point bending tests wereperformed on locking screws fordetermining their yield points with 3 different dimensions (4.5 mm, 5 mm and5,5 mm) and with 2 different screw thread (low threaded and unthreaded).Results: The mean yield point valuesof smooth 4.5 mm and threaded 5 mm low locking screws were statisticallysignificant less than that of smooth 5 mm, low threaded 5.5 mm and smooth 5.5mm,(P=0.000).Conclusions: To avoid proximal locking screw deformation,use of smooth 4,5 mm and low threaded 5 mm locking screws should be avoided innailing of comminuted tibia fracturesof unreliable persons. Smooth 5 mm, low threaded5.5 mm and smooth screws 5.5 mm may be used safely.
Amaç: Parçalı tibia kırıklarınnıntedavisinde kullanılan intramedüller çivinin kilitleme vidalarında deformasyonsıklıkla görülmektedir. Bu çalışmanın amacı tibia intramedüller çivilemesisteminde beş farklı proksimal kilitleme vidasının bükülme dirençlerinikarşılaştırmaktır.Metod: 50 adet vida üç farklıçapta(4.5 mm, 5 mm ,5.5 mm) ve 2 farklı yiv şeklinde( yivli ve yivsiz)olacakşekilde beş farklı gruba ayrıldı. 34 mm iç çapa sahip bir paslanmaz çelik tüp tibiayıtemsil etmek üzere hazırlandı. Tüm vidalara uygulama sonrası üç nokta bükülmetesti yapıldı. Sonuçlar: Yivsiz 4.5 mm ve yivli 5mmvida gruplarının ortalama değerleri yivsiz 5 mm, yivli 5.5 mm ve yivsiz 5.5 mmgruplarına gore anlamlı derecede düşük bulundu(p=0.000)Çıkarımlar: Proksimal kilitleme vidasınındeformasyonundan kaçınmak için yivsiz 4.5 mm ve yivli 5 mm vidalarınkullanılması uygun olmayabilir. Yivsiz 5 mm, yivli 5.5 mm ve yivsiz 5.5 mmvidaların güvenli bir şekilde kullanılması önerilebilir.